EA50H
0.98 T @ 20 MM - 2.37 T @ 5 MM - 27 Kg
Compact and lightweight laboratory H-electromagnet with interchangeable poles.
Specifications
Weight | 27 Kg |
Max Field | 2.37 T @ Entrefer 5mm |
Distance between coils | 43 mm |
Max power | 500 W |
Water flow | 3 L/min |
Resistance @ 20°C | 1.050 Ohm |
Inductance | 230 mH |
Documents
CAYLAR - EA50H Datasheet v1 EN
pdf . 232.24 Ko
Options
- Custom pole set (cylindrical, conical, drilled, etc.)
- 45° support
- Motorized rotation base (Manual/Motorized/Vertical or Horizontal)
- Custom coils
- Custom pole set
- AC modulation coils
Alimentation
Power supplies for the standard coil set (0.166 Ohm) :
- Monopolar power supply 20A / 25V / 0.5kW
- Options : 10ppm / 5ppm / 1ppm / Integrated HALL Gaussmeter / 4 quadrants true bipolar
Dimensions
Field and Pole Homogeneity
Gap G | B max | Face D | Homogeneity/Volume |
5 mm | 2.30 T | 12 mm | 0.90%/3x3x3 mm |
10 mm | 1.59 T | 20 mm | 1.97%/5x5x5 mm |
15 mm | 1.16 T | 22 mm | 1.82%/5x5x5 mm |
20mm | 0.90 T | 25 mm | 1.73%/5x5x5 mm |
25 mm | 0.73 T | 30 mm | 1.27%/5x5x5 mm |
30 mm | 0.61 T | 38 mm | 0.85%/5x5x5 mm |
35 mm | 0.54 T | 40 mm | 0.88%/5x5x5 mm |
40 mm | 0.47 T | 45 mm | 0.75%/5x5x5 mm |
Prices
Please find below the indicative price list to help you estimate the cost of your system. The actual price depends on a number of other factors.
Electromagnets
Set of Poles
Magnet Power Supply
Alimentation Bipolare 36V 30A 50ppm
4 Quadrants True Bipolar Power Supply
MPU-36V30A-4Q
8 500,00€
HALL Thermalized Option
Integreted HALL Gaussmeter + Thermalized HALL Probe
MPU-HALL-TH
5 000,00€
Accessories
Product added to the quote.
Specific request
Our means of simulation and our know-how allow us to manufacture a custom-made electromagnet according to 5 criteria:
- Air gap
- Maximum field
- Homogeneity
- Speed of field variation
- Mechanical constraints (Optical access, UHV, size, mass, etc.)
CAYLAR designs and manufactures magnetizing coils, vacuum chambers and current sources which offers a great flexibility for the design of an electromagnet. It is possible to adapt the coils to an existing power supply.
* Fields marked with an asterisk are required.
Data from our simulation tool. Our simulator has been calibrated to obtain values of magnetic field and homogeneity lower than the reality with an error rate between 0 and 5%.