MOKE UHV ALBA
2.28 T @ 10 MM - 0.97 T @ 40 MM - 313 Kg
Compact and lightweight laboratory electromagnet with drilled variable poles. Optimized for magneto-optical experiments (MOKE, XMCD) under ultra-high vacuum. The coils are optimized for fast ramping.
Specifications
Weight | 385 Kg |
Max Field | 2.22 T |
Distance between coils | >270 mm |
Temp. Max | 45°C |
Water flow | 12 L/min |
Resistance @ 20°C | 0.08 Ohm |
Inductance | 30 mH |
Documents
MOKE UHV ALBA - Datasheet
pdf . 248.29 Ko
Options
- Custom pole set (cylindrical, tapered, drilled, etc.)
- Custom Coils / Poles
- Modulation coils
Power supplies
Power supplies for the standard coil set:
- Power supply true Bipolar 4 quadrands 210A/40V
- Power supply true Bipolar 4 quadrands 180A/40V
- Power supply true Bipolar 4 quadrands 150A/15V
Dimensions
Field and Pole Homogeneity
Gap | B max | Face D | Homogeneity/Volume |
10 mm | 2.22T | 30 mm | 2.24%/3x3x3 mm |
11.5 mm | 2.12T | 30 mm | 2.27%/4x4x4 mm |
13 mm | 2.04 T | 30 mm | 2.40%/5x5x5 mm |
15 mm | 1.91 T | 30 mm | 1.85%/6x6x6mm |
25 mm | 1.37 T | 30 mm | 1.57%/6x6x6 mm |
30 mm | 1.16 T | 30 mm | 1.98%/6x6x6 mm |
35 mm | 1.02 T | 30 mm | 1.93%/6x6x6 mm |
40 mm | 0.89 T | 30 mm1 | 1.82%/6x6x6 mm |
Specific request
Our means of simulation and our know-how allow us to manufacture a custom-made electromagnet according to 5 criteria:
- Air gap
- Maximum field
- Homogeneity
- Speed of field variation
- Mechanical constraints (Optical access, UHV, size, mass, etc)
CAYLAR designs and manufactures magnetizing coils, vacuum chambers and current sources which offers a great flexibility for the design of an electromagnet. It is possible to adapt the coils to an existing power supply.
* Fields marked with an asterisk are required.
Data from our simulation tool. Our simulator has been calibrated to obtain values of magnetic field and homogeneity lower than the reality with an error rate between 0 and 5%.